This glossary provides clear explanations of standardized symbols used on medical device labels, packaging, and documentation. These symbols are essential for ensuring safety, compliance, and effective communication across global markets.

Symbol Description of symbol Reference number and title of symbol Title and designation number of standard
Manufacturer Indicates the medical device manufacturer, as defined in EU Directives 90/385/EEC, 93/42/EEC and 98/79/EC. 5.1.1 Manufacturer ISO 15223-1:2016(E): Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements
EC REP Indicates the authorized representative in the European Community. 5.1.2 Authorized representative in the European Community ISO 15223-1:2016(E): Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements
Date of Manufacture Indicates the date when the medical device was manufactured. 5.1.3 Date of Manufacture ISO 15223-1:2016(E): Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements
Use-by date Indicates the date after which the medical device is not to be used. 5.1.4 Use-by date ISO 15223-1:2016(E): Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements
Batch Code Indicates the manufacturer's batch code so that the batch or lot can be identified. 5.1.5 Batch code ISO 15223-1:2016(E): Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements
REF Indicates the manufacturer's catalogue number so that the medical device can be identified. 5.1.6 Catalogue number ISO 15223-1:2016(E): Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements
Sterilized using ethylene oxide Indicates a medical device that has been sterilized using ethylene oxide. 5.2.3 Sterilized using ethylene oxide ISO 15223-1:2016(E): Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements
Sterilized using irradiation Indicates a medical device that has been sterilized using irradiation. 5.2.4 Sterilized using irradiation ISO 15223-1:2016(E): Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements
Do not use if package is damaged Indicates a medical device that should not be used if the package has been damaged or opened. 5.2.8 Do not use if package is damaged ISO 15223-1:2016(E): Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements
Sterile fluid path Indicates the presence of a sterile fluid path within the medical device in cases when other parts of the medical device, including the exterior, might not be supplied sterile. 5.2.9 Sterile fluid path ISO 15223-1:2016(E): Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements
Fragile, handle with care Indicates a medical device that can be broken or damaged if not handled carefully. 5.3.1 Fragile, handle with care ISO 15223-1:2016(E): Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements
Fragile, handle with care To indicate that the contents of the transport package are fragile and the package shall be handled with care. Reference No. 0621 Fragile- handle with care ISO 7000 Graphical symbols for use on equipment - Registered symbols
Keep away from sunlight Indicates a medical device that needs protection from light sources. 5.3.2 Keep away from sunlight ISO 15223-1:2016(E): Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements
Temperature limit Indicates the temperature limits to which the medical device can be safely exposed. 5.3.7 Temperature limit ISO 15223-1:2016(E): Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements
Temperature limit To indicate the maximum and minimum temperature limits at which the item shall be stored, transported or used. Reference No. 0632 Temperature limit ISO 7000 Graphical symbols for use on equipment - Registered symbols
Humidity limitation Indicates the range of humidity to which the medical device can be safely exposed. 5.3.8 Humidity limitation ISO 15223-1:2016(E): Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements
Do not re-use Indicates a medical device that is intended for one use, or for use on a single patient during a single procedure. 5.4.2 Do not re-use ISO 15223-1:2016(E): Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements
Consult instructions for use Indicates the need for the user to consult the instructions for use.

5.4.3 Consult instructions for use

ISO 15223-1:2016(E): Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements
Caution Indicates the need for the user to consult the instructions for use for important cautionary information such as warnings and precautions that cannot, for a variety of reasons, be presented on the medical device itself. 5.4.4 Caution ISO 15223-1:2016(E): Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements
Pyrogenic Indicates a medical device that is non-pyrogenic. 5.6.3 Non-pyrogenic ISO 15223-1:2016(E): Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements
Drops per millilitre Indicates the number of drops per millilitre. 5.6.4 Drops per millilitre ISO 15223-1:2016(E): Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements
Reference No. M002 Refer to instruction manual/booklet To signify that the instruction manual / booklet must be read. Reference No. M002 Refer to instruction manual/booklet ISO 7010 Graphical Symbols – Safety colours and safety signs – Registered safety signs
Reference No. 5840 Type B applied part To identify a type B applied part complying with IEC 60601-1. Reference No. 5840 Type B applied part IEC 60417 Graphical Symbols for Use on Equipment
Reference No. 5333 Type BF applied part To identify a type BF applied part complying with IEC 60601-1. Reference No. 5333 Type BF applied part IEC 60417 Graphical Symbols for Use on Equipment
Reference No. 5007 ”ON” (power) To indicate connection to the mains, at least for mains switches or their positions, and all those cases where safety is involved. Reference No. 5007 ”ON” (power) IEC 60417 Graphical Symbols for Use on Equipment
Reference No. 5008 ”OFF” (power) To indicate disconnection from the mains, at least for mains switches or their positions, and all those cases where safety is involved. Reference No. 5008 ”OFF” (power) IEC 60417 Graphical Symbols for Use on Equipment
Reference No. 5009 Stand-by To identify the switch or switch position by means of which part of the equipment is switched on in order to bring it into the stand-by condition, and to identify the control to shift to or to indicate the state of low power consumption. Each of different states of power consumption may be indicated using a corresponding colour Reference No. 5009 Stand-by IEC 60417 Graphical Symbols for Use on Equipment
Reference No. 5032 Alternating current To indicate on the rating plate that the equipment is suitable for alternating current only; to identify relevant terminals. Reference No. 5032 Alternating current IEC 60417 Graphical Symbols for Use on Equipment
Reference No. 5019 Protective earth; protective ground To identify any terminal which is intended for connection to an external conductor for protection against electric shock in case of a fault, or the terminal of a protective earth (ground) electrode. Reference No. 5019 Protective earth; protective ground IEC 60417 Graphical Symbols for Use on Equipment
Reference No. 5021 Equipotentiality To identify the terminals which, when connected together, bring the various parts of an equipment or of a system to the same potential, not necessarily being the earth (ground) potential, e.g. for local bonding. Reference No. 5021 Equipotentiality IEC 60417 Graphical Symbols for Use on Equipment
Reference No. 0623 This way up To indicate correct upright position of the transport package Reference No. 0623 This way up ISO 7000 Graphical symbols for use on equipment - Registered symbols
Reference No. 0622 Use no hooks To indicate that hooks shall not be used for handling the transport package. Reference No. 0622 Use no hooks ISO 7000 Graphical symbols for use on equipment - Registered symbols
Reference No. 0626 Keep away from rain To indicate that the transport package shall be kept away from rain and in dry conditions Reference No. 0626 Keep away from rain ISO 7000 Graphical symbols for use on equipment - Registered symbols
Ingress Protection Against Dripping Water (vertically falling drops) To identify category AP equipment complying with IEC 60601-1 which also specifies the way in which this symbol has to be used. Ingress Protection Against Dripping Water (vertically falling drops) IEC 60601-1 Medical Electrical Equipment, Part 1: General Requirements for Safety IP degrees in accordance with IEC 60529 / EN 60529
Reference No. 2405 Do not roll To indicate that the transport package shall not be rolled or turned over but shall remain in the upright position. Reference No. 2405 Do not roll ISO 7000 Graphical symbols for use on equipment - Registered symbols
Reference No. 2402 Do not stack To indicate that the items shall not be vertically stacked, either because of the nature of the transport packaging or because of the nature of the items themselves. Reference No. 2402 Do not stack ISO 7000 Graphical symbols for use on equipment - Registered symbols
Reference No. 1321A Mass; weight To indicate mass. To identify a function related to mass. Reference No. 1321A Mass; weight ISO 7000 Graphical symbols for use on equipment - Registered symbols
Reference 5.1.7 Serial number Indicates the manufacturer’s serial number so that a specific medical device can be identified. Reference 5.1.7 Serial number ISO 15223-1 Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements
Reference No. 2403 Stacking limit by number To indicate that the items shall not be vertically stacked beyond the specified number, either because of the nature of the transport packaging or because of the nature of the items themselves. Reference No. 2403 Stacking limit by number ISO 7000 Graphical symbols for use on equipment - Registered symbols
Reference No. 5926 Polarity of d.c. power connector To identify the positive and negative connections (the polarity) of a d.c. power supply, or the positive and negative connections on a piece of equipment to which a d.c. power supply may be connected. Reference No. 5926 Polarity of d.c. power connector IEC 60417 — Graphical Symbols for Use on Equipment
Reference No. 5031 Direct current To indicate on the rating plate that the equipment is suitable for direct current only; to identify relevant terminals. Reference No. 5031 Direct current IEC 60417 — Graphical Symbols for Use on Equipment
Reference No. 2621 Atmospheric pressure limitation To indicate the acceptable upper and lower limits of atmospheric pressure for transport and storage Reference No. 2621 Atmospheric pressure limitation ISO 7000 — Graphical symbols for use on equipment — Registered symbols

Symbols not from Standards

Symbol Description of symbol Reference number and title of symbol Title and designation number of standard
21 CFR Part 801.109 Prescription use only Caution: Federal law (USA) restricts this device to sale by or on the order of a licensed healthcare practitioner. 21 CFR Part 801.109 Prescription use only FDA Quality System Regulation – Labeling requirements
Not made with natural rubber latex Not made with natural rubber latex NA NA
Not made with di(2-ethylhexyl) phthalate Not made with di(2- ethylhexyl) phthalate NA NA
Not made with Bisphenol-A Not made with Bisphenol-A NA NA
Keep transport position when moving the device on more than a 5-degree slope. Keep transport position when moving the device on more than a 5-degree slope. NA NA
Alarm sounds mute switch Alarm sounds mute switch NA NA
Blood pump switch Blood pump switch
Green – ON
Red – OFF
NA NA
Bubble detector Bubble detector NA NA
Blood sensor Blood sensor NA NA
Clamp Clamp NA NA
Not for use with Hemodiafiltration Not for use with Hemodiafiltration NA NA
This device must be used on dialysis machines with an ultrafiltration controller or accurate fluid balancing system. This device must be used on dialysis machines with an ultrafiltration controller or accurate fluid balancing system. NA NA