Product Overview
A safe, continuous supply of dialysis water is now available in a high performing, low maintenance system.
Thanks to multiple built-in safety features – including internal heat disinfection – HemoRO 4 ONE fulfills the rigorous requirements of home and acute dialysis, ensuring the utmost in patient safety.
This portable solution is ideal for dialysis centers or ICUs in need of mobility, as well as patients who wish to carry out their dialysis treatments in the comfort of home.
Benefits
- Safe: semi-automated heat disinfection (membrane, permeate path and connection hoses with >80°C hot water) ensures high water quality and microbiological safety for dialysis
- Mobile: robust housing and compact design, lightweight and easy to move thanks to built-in wheels and handle
- Easy to use: ready to use within 10 mins, low noise operation, USB and Ethernet connectivity.
|
|
Specification 230 V |
Specification 115 V |
Dimensions |
Measurements (W x H x D) |
315 x 875 x 300 mm |
315 x 875 x 300 mm |
Weight |
Dry weight |
35 kg |
35 kg |
Operating weight |
< 45 kg |
< 45 kg |
|
Performance data |
Dialysis water capacity at 10 °C and 2.5 bar counter pressure |
50 l/h A,B, 70 l/h A,C |
50 l/h A,B, 70 l/h A,C |
Dialysis water pressure |
3 bar / 0.3 MPa |
3 bar / 0.3 MPa |
|
Retention rate |
99% NaCl |
99% NaCl |
|
Noise level |
< 50 dB |
< 50 dB |
|
Water connections |
Feed water |
Hose stem D8 |
Hose stem D8 |
Dialysis water |
Nipple type 103 DN6 |
Nipple type 103 DN6 |
|
Waste water |
Hose stem D8 |
Hose stem D8 |
|
Electrical data |
Mains plug |
CEE 7/7 D |
NEMA 5-15P D |
Mains voltage |
230 V AC 50 Hz |
115 V AC 60 Hz |
|
Protective conductor cross-section area |
1.5 mm B |
18 AWG |
|
Building main fuses |
16 A, RCD IΔN = 30 mA |
B-15A acc. to UL489, RCD 30 mA type A or B |
|
Power consumption |
1.5 kW |
1.2 kW |
|
Protection class |
I |
I |
|
Protection type |
IP21 |
IP21 |
|
Feed water |
Quality |
Drinking water E |
Drinking water E |
Inflow, see possible combinations |
200 l/h B, 250 l/h C |
200 l/h B, 250 l/h C |
|
Feed water pressure, see possible combinations |
> 1.5 bar / 0.15 MPa B > 2.5 bar / 0.25 MPa C < 8.0 bar / 0.8 MPa |
> 1.5 bar / 0.15 MPa B > 2.5 bar / 0.25 MPa C < 8.0 bar / 0.8 MPa |
|
Hardness |
< 15 °dH / 2.45 mmol/l |
< 15 °dH / 2.45 mmol/l |
|
Blocking index/SDI |
< 3 |
< 3 |
|
Conductivity |
100 to 2,000 μS/cm |
100 to 2,000 μS/cm |
|
Iron |
< 0.05 mg/l |
< 0.05 mg/l |
|
Free chlorine |
< 0.1 mg/l |
< 0.1 mg/l |
|
Temperature range |
5 to 30 °C |
5 to 30 °C |
|
pH |
6.5 to 8.5 |
6.5 to 8.5 |
|
Waste water |
Free drainage |
max. 50 cm above device top edge |
max. 50 cm above device top edge |
Environmental conditions |
Transport / storage temperature |
1 to 45°C |
1 to 45°C |
Operating temperature |
5 to 35°C |
5 to 35°C |
|
Relative air humidity |
max. 90% |
max. 90% |
|
Installation altitude |
< 2,000 m above NHN F |
< 2,000 m above NHN F |
Possible combinations
HemoRO 4 ONE: ECO
|
Inflow |
Feed water pressure |
|
(50 l/h) |
> 200 l/h |
> 1,5 bar (dyn.) |
< 8,0 bar (stat.) |
+ with pre-filter |
> 200 l/h |
> 2,5 bar (dyn.) |
< 8,0 bar (stat.) |
+ with pre-filter system |
> 200 l/h |
> 4,5 bar (dyn.) |
< 8,0 bar (stat.) |
+ with softener |
> 300 l/h |
> 2,0 bar (dyn.) |
< 8,0 bar (stat.) |
+ with pre-filter system and softener |
> 300 l/h |
> 4,5 bar (dyn.) |
< 8,0 bar (stat.) |
HemoRO 4 ONE: MAX
|
Inflow |
Feed water pressure |
|
(70 l/h) |
> 250 l/h |
> 2,5 bar (dyn.) |
< 8,0 bar (stat.) |
+ with pre-filter |
> 250 l/h |
> 3,5 bar (dyn.) |
< 8,0 bar (stat.) |
+ with pre-filter system |
> 250 l/h |
> 5,5 bar (dyn.) |
< 8,0 bar (stat.) |
+ with softener |
> 350 l/h |
> 3,5 bar (dyn.) |
< 8,0 bar (stat.) |
+ with pre-filter system and softener |
> 350 l/h |
> 6,0 bar (dyn.) |
< 8,0 bar (stat.) |