Product Overview
A safe, continuous supply of dialysis water is made possible in a portable, easy to use system.
HemoRO 3000 is ideal for dialysis centers or ICU's in need of mobility, as well as patients who wish to carry out their dialysis treatments in the comfort of home.
Benefits
Semi-automated chemical disinfection and fully-automated standby flushing ensures a high water quality
- Compact design with robust metal housing and easy to move thanks to built-in wheels
- Simple access to settings via a 2-button user interface
- Integrated pre-filter with pressure reducer
|
|
Specification |
|
Dimensions |
Measurements (W x H x D) |
480 x 170 x 360 mm |
|
Weight |
Max |
35 kg |
|
Performance data |
Maximum dialysis water production capacity at 10 °C and with a back pressure of 2.5 bar |
02HRO3000 50 l/h |
|
02HRO3002 70 l/h |
|||
02HRO3110 50 l/h |
|||
02HRO3111 70 l/h |
|||
Dialysis water pressure |
3 bar |
||
Membrane retention rate |
> 95% |
||
Water connections |
Feed water |
Hose DN 8 |
|
Dialysis water |
Hose DN 6 (2x 1.5 m) |
||
Waste water |
Hose DN 8 |
||
Electrical data |
Mains plug |
230 V-Version: CEE 7/7 115 V-Version: NEMA 5-15 P |
|
Supply voltage |
230 V / 50-60 Hz or 115 V / 60 Hz |
||
Energy consumption |
max. 0.3 kW |
||
Fuse |
16 A / Fi ΔI 30 mA |
||
Electrical safety |
Safety class |
I |
|
Overvoltage category |
II |
||
Degree of pollution |
2 |
||
Feed water |
Quality |
Drinking water or softened drinking water |
|
Inflow, see possible combinations |
|
||
Hardness |
02HRO3000 up to 4.45 mmol/l 02HRO3002 up to 2.45 mmol/l 02HRO3110 up to 4.45 mmol/l 02HRO3111 up to 2.45 mmol/l |
||
SDI |
< 3 |
||
Conductivity |
< 2,000 μS/cm |
||
Iron |
< 0.05 mg/l |
||
Free chlorine |
< 0.10 mg/l |
||
Temperature range |
5 – 25°C |
||
pH-value |
6.5 – 8.5 |
||
Ambient conditions |
Temperature |
Shipping / Storage |
+1 to +45°C |
Operation |
+1 to +35°C |
||
Air humidity |
max. 90% rel. humidity, non-condensing |
||
Installation altitude |
max. 2,000 m above NHN |
||
On-site requirements |
Isolable water supply, 2.5 – 6 bar |
Pressure reducing valve or pressure booster as required |
|
Drain max. 100 cm above upper edge of the unit |
|
Possible combinations
|
Feed Flow |
Feed water pressure |
|
(50l/h) |
200 l/h |
> 2,5 bar (dyn.) |
< 10,0 bar (stat.) |
+ softener |
300 l/h |
> 3,0 bar (dyn.) |
< 8,0 bar (stat.) |
+ 3x pre-filter |
200 l/h |
> 4,0 bar (dyn.) |
< 8,5 bar (stat.) |
+ 3x pre-filters + softener |
300 l/h |
> 4,5 bar (dyn.) |
< 8,0 bar (stat.) |
|
Inflow |
Feed water pressure |
|
(70l/h) |
220 l/h |
> 2,5 bar (dyn.) |
< 10,0 bar (stat.) |
+ softener |
320 l/h |
> 3,0 bar (dyn.) |
< 8,0 bar (stat.) |
+ 3x pre-filter |
220 l/h |
> 4,0 bar (dyn.) |
< 8,5 bar (stat.) |
+ 3x pre-filters + softener |
320 l/h |
> 4,5 bar (dyn.) |
< 8,0 bar (stat.) |